3-10 AeroTrak
®
Handheld Airborne Particle Counters
The following table describes the options of this menu and the parameters
you can set.
C A U T I O N
If you turn Data Logging off, no data will be saved to the buffer.
Option
Description
Parameters
Pump Speed
Press the Enter key to go to a
secondary screen where you can
increase or decrease the pump speed.
(The pump may slow down with age,
or it may be necessary to increase the
speed if there are flow restrictions
such as long tubing). Use a flow meter
to check the flow when taking critical
measurements, and adjust the pump
speed as necessary.
0 to 255
Data Logging
Use < and > to turn data logging on
and off.
ON; OFF
Locations
Press the Enter key to go to a
secondary screen where you can
select a location number.
0 to 250