3-10 AeroTrak
®
Handheld Airborne Particle Counters
The following table describes the options of this menu and the parameters
you can set.
Press the Enter key to go to a
secondary screen where you can
increase or decrease the pump speed.
(The pump may slow down with age,
or it may be necessary to increase the
speed if there are flow restrictions –
such as long tubing). Use a flow meter
to check the flow when taking critical
measurements, and adjust the pump
speed as necessary.
Press the Enter key to go to a
secondary screen where you can
select a location number.