A-1
A P P E N D I X A
Specifications
All specifications meet or exceed ISO 21501-4 and JIS B9921. They are
subject to change without notice.
Specification
Description
Size Range ........................................
0.3 to 25 µm
Channel Sizes ...................................
Standard: 0.3, 0.5, 0.7, 1.0, 2.0, 5.0 μm
Standard: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 μm
Standard: 0.3 to 10 μm, user-selectable; factory-calibrated at 0.3, 0.5, 1.0, 3.0,
5.0, 10.0 μm. Additional channel sizes available
Size Resolution ................................
<15% at 0.5 μm (per ISO 21501-4 requirements)
Counting Efficiency ..........................
50% at 0.3 μm; 100% for particles > 0.45 μm (per JIS and ISO 21501-4)
Concentration Limits .......................
5,950,000 particles/ft
3
at 10% coincidence loss
Light Source .....................................
Long life laser diode
Zero Count Level ..............................
<1 count/5 minutes (per JIS B9921 and ISO 21501-4)
Flow Rate ..........................................
0.1 CFM (2.83 L/min) with 5% accuracy (meets JIS and ISO 21501-4
requirements)
Calibration ........................................
NIST traceable with TSI
®
calibration system
Calibration Frequency .....................
Recommended minimum once per year
Sample Probe/Tubing ......................
Isokinetic sampling probe
Sampling Modes ...............................
Manual, automatic, beep, cumulative/differential count or concentration
Sampling Time ..................................
1 second to 99 hours
Sampling Frequency ........................
1 to 9999 cycles or continuous
Exhaust .............................................
Internally filtered
Vacuum Source ................................
Internal pump
Communication Mode ......................
Modbus
®
TCP over Ethernet or USB
Data Storage .....................................
10,000 sample records: includes date, time, six particle channels, flow, ID, and
sample volume; transferable via USB data download or TrakPro™ Lite software
Data Security ....................................
Password protected
Alarm/Status .....................................
Audible alarm sounds for counts over threshold values set in recipe, low battery,
and for instrument or flow errors.
Environmental Sensors ...................
Optional temperature/RH probe supported
Display ..............................................
QVGA 3.5-inch (8.9-cm) touch screen with Windows
®
CE operating system
Languages ........................................
English, Spanish, German, French, Italian, Japanese, and Chinese (simplified)
Reports ..............................................
Provides Pass/Fail on FS-209E, ISO 14644-1 and EU GMP
Printer ................................................
Optional external printer supported

Other manuals for TSI Instruments AEROTRAK 9306

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments AEROTRAK 9306 and is the answer not in the manual?

TSI Instruments AEROTRAK 9306 Specifications

General IconGeneral
BrandTSI Instruments
ModelAEROTRAK 9306
CategoryCash Counter
LanguageEnglish

Related product manuals