A-1
A P P E N D I X A
Specifications
All specifications meet or exceed ISO 21501-4 and JIS B9921. They are
subject to change without notice.
Size Range ........................................
Channel Sizes ...................................
Standard: 0.3, 0.5, 0.7, 1.0, 2.0, 5.0 μm
Standard: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 μm
Standard: 0.3 to 10 μm, user-selectable; factory-calibrated at 0.3, 0.5, 1.0, 3.0,
5.0, 10.0 μm. Additional channel sizes available
Size Resolution ................................
<15% at 0.5 μm (per ISO 21501-4 requirements)
Counting Efficiency ..........................
50% at 0.3 μm; 100% for particles > 0.45 μm (per JIS and ISO 21501-4)
Concentration Limits .......................
5,950,000 particles/ft
3
at 10% coincidence loss
Light Source .....................................
Zero Count Level ..............................
<1 count/5 minutes (per JIS B9921 and ISO 21501-4)
Flow Rate ..........................................
0.1 CFM (2.83 L/min) with 5% accuracy (meets JIS and ISO 21501-4
requirements)
Calibration ........................................
NIST traceable with TSI
®
calibration system
Calibration Frequency .....................
Recommended minimum once per year
Sample Probe/Tubing ......................
Isokinetic sampling probe
Sampling Modes ...............................
Manual, automatic, beep, cumulative/differential count or concentration
Sampling Time ..................................
Sampling Frequency ........................
1 to 9999 cycles or continuous
Exhaust .............................................
Vacuum Source ................................
Communication Mode ......................
Modbus
®
TCP over Ethernet or USB
Data Storage .....................................
10,000 sample records: includes date, time, six particle channels, flow, ID, and
sample volume; transferable via USB data download or TrakPro™ Lite software
Data Security ....................................
Alarm/Status .....................................
Audible alarm sounds for counts over threshold values set in recipe, low battery,
and for instrument or flow errors.
Environmental Sensors ...................
Optional temperature/RH probe supported
Display ..............................................
QVGA 3.5-inch (8.9-cm) touch screen with Windows
®
CE operating system
Languages ........................................
English, Spanish, German, French, Italian, Japanese, and Chinese (simplified)
Reports ..............................................
Provides Pass/Fail on FS-209E, ISO 14644-1 and EU GMP
Printer ................................................
Optional external printer supported