Operation 3-41
Data Integrity Mode
Data Integrity Mode allows the unit to be locked down to ensure the
sampling data cannot be modified. When in this mode, all printing and
data exporting is prohibited.
Data Integrity Mode is designed to be used in conjunction with TrakPro
Lite Secure software. When enabling data integrity mode on the
instrument, only the data can be viewed.
The lock icon on the main page indicates the unit is in data integrity
mode, but no user is logged in.
To allow sampling, the unit will need to be unlocked using TrakPro Lite
Secure software. During this process TrakPro Lite Secure software will
place the user ID # and password into the instrument to allow you to
take samples.