22
Note:
l The test leads can add 0.1 to 0.2 of error to
resistance measurement. To obtain precision readings
in low-resistance measurement (UT60F: 400 and
UT60G is 600), short-circuit the input terminals
beforehand and record the reading obtained (called
this reading as X).
Then use the equation:
measured resistance value (Y) –(X) = precision
readings of resistance.
l For high-resistance measurement (>1M), it is normal
taking several seconds to obtain a stable reading.
l If reading with shorted test leads is not 0.5,
check for loose test leads, wrong function selected,
or enabled data hold function.
l The LCD displays OL indicating open-circuit or the
tested resistor value is higher than the maximum
range of the Meter.
l Resistance measurement is default to auto range
mode. At range, open circuit voltage approximate
1.20V. At all other range, open circuit voltage
approximate 0.45V.
l When resistance measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test and remove testing leads
from the input terminals.
Measurement Operation(6)
Model UT60F/G: OPERATING MANUAL