51
Model UT805: OPERATING MANUAL
C. DC Current (Calibrate both positive and negative
polarity)
2mA :
200mA :
10A :
Press REL to display “0”, input 1.9mA,
press CAL button to display (--CAL--)
(-HI-END) ( 1.90000mA), calibration
finished.
Press REL to display “0”, input 190mA,
press CAL button to display (--CAL--)
(-HI-END) ( 190.000mA), calibration
finished
Press REL to display “0”, input 10A,
press CAL button to display (--CAL--)
(-HI-END) ( 10.0000A), calibration
finished
D. AC Current (Frequency: 1kHz)
2mA :
200mA :
10A :
AC+DC:
Input 0.19mA, Press CAL button to display
(--CAL--) (-LO-END) ( ) Input
1.9mA, Press CAL button to display (--CAL--)
(-HI-END) (1.90000mA), calibration
finish.
Input 19mA, Press CAL button to display
(--CAL--) (-LO-END) ( ) Input
190mA, Press CAL button to display
(--CAL--) (-HI-END) (190.000mA),
calibration finish
Input 1.9A, Press CAL button to display
(-- CAL--) (-LO-END) ( ) Input
10A, Press CAL button to display
(-- CAL--) (-HI-END) (10.0000A),
calibration finish
AC+DC:Carry out the calibration according
to each range of DC current.