Model UT805: OPERATING MANUAL
E. Resistance F. AC Voltage with DC deviation (AC+DC)
1. Press ACV button
2. Press AC+DC button
3. Follow the above “B. AC Voltage” procedure
G. AC Current with DC deviation (AC+DC)
1. Press ACI button
2. Press AC+DC button
3. Follow the above “D. AC Current” procedure.
2k :
20k :
200k :
2M :
20M :
Press REL to display “0”, input 1.9k ,
press CAL button to display (--CAL--)
(-HI-END) ( 1.90000k ), calibration
finished.
Press REL to display “0”, input 19k ,
press CAL button to display (--CAL--)
(-HI-END) ( 19.0000k ), calibration
finished.
Press REL to display “0”, input 190k ,
press CAL button to display (--CAL--)
(-HI-END) ( 190.000k ), calibration
finished.
Press REL to display “0”, input 1.9M ,
press CAL button to display (--CAL--)
(-HI-END) ( 1.90000M ), calibration
finished.
Press REL to display “0”, input 19M ,
press CAL button to display (--CAL--)
(-HI-END) ( 19.0000M ), calibration
finished