User’s Manual UTG1000X Series
Instruments.uni-trend.com 16 / 74
Chapter 2 Quick Start
3.1 Output Fundamental Wave
3.1.1 Output Frequency
The default waveform is a sine wave with frequency 1 kHz, amplitude 100 mV peak-to-peak (connect with
50Ω port).
The specific steps to change the frequency to 2.5 MHz,
Press WaveSineFrequency key in turn, use numerical keyboard to input 2.5 and then select the unit of
the parameter to MHz.
3.1.2 Output Amplitude
The default waveform is a sine wave with amplitude 100 mV peak-to-peak (connect with 50Ω port).
The specific steps to change the amplitude to 300mVpp,
Press Wave SineAmp key in turn, use numerical keyboard to input 300 and then select the unit of the
parameter to mVpp.
3.1.3 DC Offset Voltage
DC offset voltage of the waveform is 0V sine wave in default (connect with 50Ω port).
The specific steps to change DC offset voltage to -150mV,
Press Wave SineOffset key in turn, use numerical keyboard to input -150 and then select the unit of the
parameter to mVpp.
Notes: Multifunction and arrow key can also be used to set the parameter.
3.1.4 Phase
The phase of the waveform is 0° in default.
The specific steps to set the phase to 90°,
Press Phase key, use numerical keyboard to input 90 and then select the unit of the parameter to °.
3.1.5 Duty Cycle of Pulse wave
The default frequency of pulse wave is 1 kHz, duty cycle 50%.
The specific steps to set duty cycle to 25% (limited by a minimum pulse width specification of 22ns),
Press Wave pulseDuty key in turn, use numerical keyboard to input 25 and then select the unit of the
parameter to % .
3.1.6 Symmetry of Ramp Wave
The default frequency of ramp wave is 1 kHz, take triangular wave with the symmetry 75% as an example,