Damaged
Strip Inserted
Remove the test strip
and shut off. Turn on
and repeat test with new
strip. Confirm there are
no tears, wrinkles or
debris.
Optical detec-
tor needs cle-
aning
Remove strip holder
and clean lenses follo-
wing the direction.
Optical detec-
tor does not
read the back-
ground blank
Before finishing the ba-
ckground blank reading,
do not apply blood sam-
ple. Do it until displayi-
ng a flashing prompt.
E4
Optical detec-
tor is clogged
Check the Optical dete-
ctor, whether it is cove-
red by blood or others,
if yes, remove the strip
holder and clean it. Ref-