Decontamination by autoclaving
121°C
2 bar absolute
Autoclaving
Thecollectionbottlewithbottleheadandscrewcap,thequick
coupling,thelterandtheVHC
pro
with adapters and aspirati-
onhosearedesignatedforsteamsterilizationat121°Cand
2barabsolute(1baroverpressure).Timeofexposureaccor-
ding to DIN 58946 t
e
= 20 minutes.
Ensurenottoexceedatemperatureof121°C.
Comply with instructions for use of the manufacturer of the
autoclave.
It is the user´s responsibility to ensure effective autoclaving.
Prior to autoclaving loosen or remove the bottle head
from the bottle.
M
i
d
i
s
a
r
t
®
2
0
0
0
0
.
2
0
µ
m
P
T
F
E
IN
sartorius stedim
Prior to autoclaving set the VHC
pro
to “continuous aspira-
tion / autoclaving”.
The applicability of the decontamination method has to be ensured by the user.
Priortodecontaminationcleansurfaceswithadampclothifnecessary.Usewateror
a mild detergent as cleaning agent.
NOTICE
NOTICE
NOTICE

Other manuals for vacuubrand BVC basic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the vacuubrand BVC basic and is the answer not in the manual?

vacuubrand BVC basic Specifications

General IconGeneral
Brandvacuubrand
ModelBVC basic
CategoryWater Pump
LanguageEnglish

Related product manuals