page 17 of 27
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 25/11/2013 // 999281
Tools required (metric):
- DiaphragmkeySW46(orderno.:637777)
- ScrewdriverTorxTX20
- Open-endedwrenchSW19
Decontamination of the pump parts in a disinfectant solution or by autoclaving
The vacuum pump has to be disassembled.
When disassembling avoid contact of contaminated parts with not contami-
nated parts.
The pictured pump parts have to be decontaminated.
The applicability of the decontamination method and the selection of the ap-
propriate disinfectant have to be ensured by the user.
Priortodecontaminationcleansurfaceswithadampclothifnecessary.Usewateror
a mild detergent as cleaning agent.
NOTICE
NOTICE
NOTICE