CHAPTER 4_______________________________________________ SERIAL COMMUNICATION SETTINGS
VAISALA _________________________________________________________________________ 13
MF_Rx_LOW yyy<cr>
where:
x = 1 or 2 (number of the relay)
yyy = trigger point (CO
2
content in ppm)
Example of setting the lower trigger limit of the relay 1:
>MF_R1_Low 900
MF_R1_LOW=900.000000
>
NOTE
Remember to save the settings.
Pressure compensation
GMT220 series transmitters have pressure compensation in the
software calculations. The factory set value for current pressure is
1013 hPa. For example in case of high altitude, the pressure value can
be changed by using the COM adapter (19040GM), a computer and a
suitable terminal software. The pressure is given in hPa.
MF_PRESSURE xxxx<cr>
where:
xxxx = pressure in hPa
Example of setting desired pressure:
>MF_PRESSURE 900
MF_PRESSURE=900
>
NOTE
Remember to save the settings.
Saving the settings
MAIN_SAVE F<cr>
All the above mentioned settings have to be saved by using the
command MAIN_SAVE F<cr>.