OUTPUT MODIFICATION FIELDS
[text field]
Type the text you want inside the quotation marks "text".
[/]
Type / to return the factory setting of FORM command.
Only alone, without any other specifiers.
[ASCII character]
Type \xxx where xxx is a three digit decimal number of the
character, for example \035 outputs the # character.
[line feed]
Type \n
[carriage return]
Type \r
[space]
Type " " (Note! Leave a space in between the quotation
marks)
[horizontal tabulation]
Type \t
If you have problems in typing \ (backslash) character, you can use #
instead of it.
Use \\ for literal backslash and ## for literal hash.
Examples:
1. Output needed: Dewpoint/Frostpoint temperature with a unit
label by using three digits before and after the point. The error
value shown after the reading. Carriage return and line feed in
the end.
>form "Tdf=" 3.3 tdf U +tdf U \r \n
"Tdf=" 3.3 Tdf U +Tdf U \r \n
>send
Tdf= 0.383'C 0.087'C
>
2. Output needed: Dewpoint/Frostpoint temperature and the error
with a unit label. Time and status to be shown. The fields are
separated with horizontal tabulation. Carriage return and line
feed in the end.
>form "Tdf="3.3 tdf u \t +tdf u \t time \t status
\r \n
"Tdf=" 3.3 Tdf U \t +Tdf U \t TIME \t STATUS \r \n
>send
Tdf= 0.860'C 0.106'C 15:30:26 OK
>