EasyManua.ls Logo

Vaisala DEWCAP DM500 - TIME Setting the Date and Time; RSEL Setting the Relays

Vaisala DEWCAP DM500
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User's Guide ______________________________________________________________________
96 __________________________________________________________________ M210205EN-D
NOTE
When you want to have both metric and non-metric temperature units
(C/F) for different quantities, select the units with keypad. The serial
command UNIT changes all temperature units either to metric or
non-metric.
Examples:
>unit ?
Output units : 'C g/m3 gr/lb kJ/kg bar slpm
Selecting non-metric units
>unit n
Output units : 'F gr/ft3 gr/lb Btu/lb psi scfh
Changing the pressure unit
>unit bar
Output units : 'F gr/ft3 gr/lb Btu/lb bar scfh
TIME Setting the Date and Time
Syntax: TIME [date time]<cr>
where
date = yyyy-mm-dd (24 hour clock)
time = hh:mm:ss
Example:
>time
Current time: 2000-01-01 02:55:17
Enter new date (yyyy-mm-dd): 2001-10-05
Enter new time (hh:mm:ss): 12:07:00
RSEL Setting the Relays
Syntax: RSEL [q1 q2]<cr>
where
q1 = quantity for the relay 1
q2 = quantity for the relay 2

Table of Contents