Errors
Cause and Possible Solution
HTM10 error Problem with HTM10 module.
- Check that the module sits firmly
in place. Remove and reconnect.
- Check for missing or damaged
HUMICAPĀ® sensor.
- Check for condensation on the
HUMICAPĀ® sensor. Wait for the
sensor to dry out.
- Replace the module if unable to
remove the problem.
GM10 error Problem with GM10 module.
- Check that the module sits firmly
in place. Remove and reconnect.
- Check that supply voltage is in
range.
- Replace the module if unable to
remove the problem.
Internal error Internal problem with the transmitter.
- Restart the transmitter.
- Restore the factory settings using
service port if reset does not
help.
- Contact helpdesk@vaisala.com if
unable to remove the problem.