Syntax Description
asel ?<cr>
Show currently set analog output
parameters and scaling.
asel<cr>
Show currently set analog output
parameters and scaling, prompt to enter new
scaling values.
asel [ch1 ch2]<cr>
Set new output parameters for both
channels, prompt to enter new scaling
values. Selectable parameters for ch1 and
ch2 are:
t = temperature
rh = relative humidity
td = dewpoint
h = enthalpy
tw = wet bulb temperature
asel [ch1ch2
[ch1loch1hich2loch2hi]]<cr>
Set analog output parameters and scaling for
both channels without prompting.
ch1 = Output parameter for channel 1.
Parameters as above.
ch2 = Output parameter for channel 2.
Parameters as above.
ch1lo = Low limit for channel 1 scaling.
ch1hi = High limit for channel 1 scaling.
ch2lo = Low limit for channel 2 scaling.
ch2hi =High limit for channel 2 scaling.
Example (show current analog output parameters and scaling):
asel ?
Ch1 RH lo : 0.00 %RH
Ch1 RH hi : 100.00 %RH
Ch2 T lo : -40.00 'C
Ch2 T hi : 60.00 'C
Table 27 Asel Command