Syntax Description
crhclr<cr>
Clear the current user adjustment for
humidity. Factory calibration remains.
Example:
crhclr
OK
Table 30 Crhclr Command
Syntax Description
ct<cr>
Start the two-point temperature calibration
and adjustment sequence. For a full
adjustment procedure, see section Two-
Point Temperature Calibration and
Adjustment using a Computer on page49.
ct [ref]<cr>
Perform a one-point adjustment at the
current temperature. The device calculates
the required correction based on the
reference temperature that you enter.
ref =the correct temperature (in degrees
Celcius) at the environment where the
sensor is now.
Example :
ct
T : 22.9424 1. ref ? 23.0
Press any key when ready ...
T : 54.9873 2. ref ? 55
OK
Table 31 Ct Command
Syntax Description
ctclr<cr>
Clear the current user adjustment for
temperature. Factory calibration remains.
Example:
ctclr
OK
Table 32 Ctclr Command