Chapter 8 ____________________________________________ Sensor and Data Message Settings
VAISALA______________________________________________________________________ 143
Changing the auto-send mode to the tipping bucket mode:
0RU,M=C<cr><lf>
0RU,M=C<cr><lf>
Making the Rain amount Rc and Rain intensity Ri available both in the
precipitation message and composite data message:
0RU,R=1010000010100000<cr><lf>
0RU,R=10100000&10100000<cr><lf>
The response after the change:
0R3<cr><lf>
0R3,Rc=0.00M,Ri=0.0M<cr><lf>
Example (SDI-12, device address 0):
Changing the counter reset mode (resets the precipitation counters):
0XRU,Z=M!0<cr><lf>
In SDI-12 mode a separate enquiry (0XRU!) must be given to check the
data content.