Sensors
3.5 Optical Sensors
An optical sensor must be kept clean to operate correctly. Ensure that the SWiFT plus is powered
down before cleaning the sensor.
Use warm soapy water with a soft bristled brush to remove any light fouling.
For heavy fouling use a solvent (e.g Isopropyl alcohol) and a soft bristled brush.
Always rinse thoroughly after every use in clean, fresh water.
3.5.1 Fluorometers
3.5.1.1 Safety Statement
A Valeport Fluorometer is classified as Risk Group 1 under standard 62471. As the type is
classified as Risk Group 1 solely due to radiation in the visible band a hazard label is not required.
However,
the LED used is in excess of the Exempt Group and that the viewer-related risk is
dependent upon how the user installs and operates the equipment.
The exposure hazard value (EHV) for a Valeport Fluorometer in terms of distance is
320mm.
Never look directly into the optical aperture
3.5.1.2 Chlorophyll a
Performance
Excitation:
470 nm
Detection:
696 nm
Dynamic Range:
0 - 800 µg/l
Instrument Detection limit:
0.025 µg/l*
Actual Detection limit:
0.025 µg /l**
Linearity:
0.99 R
2
Response Time:
0.03 to 2 sec
* 3x SD in RO water
** calibrated against Chlorophyll a in acetone solution
3.5.1.3 Fluorescein
Performance
Excitation:
470 nm
Detection:
545 nm
Dynamic Range:
0 - 500 ppb
Instrument Detection limit:
<0.01 ppb*
Actual Detection limit:
0.03 ppb**
Linearity:
0.99 R
2
Response Time:
0.03 to 2 sec
* 3x SD in RO water
** Calibrated against Fluorescein solution