EasyManuals Logo

Vatech A9 User Manual

Vatech A9
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #191 background imageLoading...
Page #191 background image
15. Appendices
vatech A9 (Model name: PHT-30CSS) User Manual 191
English
15.5.2 Forward head posture vs Normal posture: X-ray path
When the patient is in a normal posture, the "distance of the cervical region through
which the X-ray passes" is short (orange arrow), so attenuation is small, but when the
patient is in the Forward head posture, the "distance of the cervical region through
which the X-ray passes" is relatively long Because of this (red arrow), the attenuation
is large. Therefore, it may occur as artifacts in the image depending on the degree of
attenuation.
Forward head posture Normal C-shaped neck posture
X-ray
path
Generator
Detector
Generator
Detector

Table of Contents

Other manuals for Vatech A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech A9 and is the answer not in the manual?

Vatech A9 Specifications

General IconGeneral
BrandVatech
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals