3. Warnings and Precautions
vatech A9 (Model name: PHT-30CSS) User Manual 27
English
Lasers
The system incorporates Class 1 laser products. The light localizers used in this
product are intended for correct patient positioning and must not be used for any
other purpose.
For maximum safety, advise the patient not to look directly at the laser beam.
While adjusting the patient, ensure that the laser beam is not directed at the
patient’s eyes.
Wavelength: 650 nm, Radiant power: Max. 039 mW
Risk of eye injury!
Do not use this equipment with any other laser sources and
do not make any changes to the settings or processes that
are described in these operating instructions.
Cleaning
Never expose this equipment to liquids, mists, or sprays. Exposing this equipment
to liquids may cause an electric shock or otherwise damage the system.
Do not use spray cleaners on this equipment, as this could cause a fire.
During the Operation
Never use this equipment in an environment that is susceptible to explosion.
Do not place flammable materials near this equipment.
Do not operate the PC while the equipment is operating. Failure to comply with
this instruction may result in system malfunction.
Immediately stop imaging if the equipment malfunctions in any way.
If a problem occurs during imaging, press the red Emergency Stop Switch to
immediately stop all moving parts and cut off all power to the equipment’s
electrical components.
Never touch the patient while the operator is touching the SIP/SOP connectors.
The medical electrical equipment or medical electrical system should not be used
adjacent to or stacked with other equipment and that if adjacent or stacked use is
necessary, the medical electrical equipment or medical electrical system should
be observed to verify normal operation in the configuration in which it will be
used.
The use of accessories and cables other than those specified, except cables sold
by VATECH of the medical electrical equipment or medical electrical system as
replacement parts for internal components, may result in increased EMISSIONS
or decreased IMMUNITY of EQUIPMENT or SYSTEM.

Table of Contents

Other manuals for Vatech A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech A9 and is the answer not in the manual?

Vatech A9 Specifications

General IconGeneral
BrandVatech
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals