EasyManuals Logo

Vatech PHT-30CSS User Manual

Vatech PHT-30CSS
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #122 background imageLoading...
Page #122 background image
10. Troubleshooting
122 vatech A9 (Model name: PHT-30CSS) User Manual
10.2.2 Main code Software (002)
10.2.2.1 Subcode – Sequence related error (001)
Tail
code
Description
001 Appears when the packing mode is enabled
002 Appears when the door is open
003 Appears when the exposure switch is pressed
10.2.2.2 Subcode – PC Resolution related error (010)
Tail
code
Description
001 Appears when the resolution is less than 1280x1024
002 Appears when the resolution is less than 1200x960
10.2.2.3 Subcode – PC Network related error (024)
Tail
code
Description
002 Appears when the port is invalid
003 Appears when the time is out
10.2.3 Main code - Acquisition Module (003)
10.2.3.1 Subcode – Initialization Failure related error (010)
Tail
code
Description
000 Appears when the COM port cannot be opened
001
Appears when the frame grabber interface cannot be initialized, or
memory for acquisition cannot be reserved
002
Appears when the MCU is not communicable, or the modem ring signal
is in an improper state

Table of Contents

Other manuals for Vatech PHT-30CSS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech PHT-30CSS and is the answer not in the manual?

Vatech PHT-30CSS Specifications

General IconGeneral
BrandVatech
ModelPHT-30CSS
CategoryMedical Equipment
LanguageEnglish

Related product manuals