EasyManuals Logo

Vatech PHT-30CSS User Manual

Vatech PHT-30CSS
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #128 background imageLoading...
Page #128 background image
12. Maintenance
128 vatech A9 (Model name: PHT-30CSS) User Manual
12.2 Maintenance Task Checklist
Tasks Period
Before the operation, ensure that the equipment is clean and ready for
use. Make sure that all parts that come in direct contact with the patient
have been cleaned thoroughly.
Daily
After using the equipment, make sure that the Main Power Switch has
been turned off.
Daily
Ensure that the equipment is firmly plugged into a dedicated power
source.
Daily
Ensure that the plug and the power cord are not heated abnormally. Daily
Confirm that the LED indicator turns yellow when the Exposure Switch
is pressed. Ensure that the LED indicator remains yellow for the entire
duration of the exposure.
Daily
Ensure that the power cable is not kinked, broken, exposed, and free of
all other defects.
Daily
Confirm that activating the Emergency Stop Switch ceases the unit’s
operation. Pressing the Emergency Stop Switch should stop all
movement of the equipment and X-ray emission.
Weekly
Ensure that all visible labels are intact and legible. Weekly
Check for damages to the Exposure Switch cable. Monthly
Confirm that the audio message is audible throughout the exposure. Monthly

Table of Contents

Other manuals for Vatech PHT-30CSS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech PHT-30CSS and is the answer not in the manual?

Vatech PHT-30CSS Specifications

General IconGeneral
BrandVatech
ModelPHT-30CSS
CategoryMedical Equipment
LanguageEnglish

Related product manuals