EasyManua.ls Logo

Vatech PHT-60CFO

Vatech PHT-60CFO
167 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PaX-i3D Green (PHT-60CFO) User Manual
153
Appendices
2.3
X-ray Scatter Dose
Test Equipment Information
Instrument Manufacturer Model S / N
Dose Meter Radcal 9015 91-1431/9913,4
PANO Mode (Xmaru1501CF)
Test Condition
Mode Tested
PANO HD / Adult
Applied Tube Voltage Peak [kVp]
80
Applied Tube Current [mA]
16
PANO HD 13.5 s [mR/hr]
1 m
(3.3 ft)
1.5 m
(4.9 ft)
2 m
(6.6 ft)
0
0.070 0.053 0.041
45
0.074 0.055 0.039
90
0.066 0.055 0.045
135
0.325 0.170 0.089
180
0.374 0.162 0.071
225
0.251 0.171 0.117
270
0.193 0.106 0.058
315
0.085 0.072 0.060
Mode
Direction [°]
[PI3DG_130U_44A_en]User Guide.indd 153 2016-05-24 오후 4:19:23

Table of Contents

Related product manuals