7 Set up with the display and adjustment module
POINTRAC 31 • 8/16 mA/HART - four-wire
43388-EN-180322
Keep the entered PIN number in mind. Operation of the sensor is only
possible with this PIN number.
Caution:
When the sensor is locked, adjustment via PACTware/DTM as well as
other systems is also blocked.
In delivery status, the PIN is 0000.
Callourservicedepartmentifyouhavemodiedandforgottenthe
PIN.
Afterachange,allsafety-relevantparametersmustbeveried.For
this purpose, a character string comparison must be carried out. This
is used to check the character respresentation and the communica-
tion channels.
Conrmifthetwocharacterstringsareidentical.Thevericationtexts
are provided in German and in the case of all other menu languages,
in English.
Inasecondstep,allmodiedsafety-relevantparametersarelisted.
Conrmthemodiedvalues.
If the described process of parameter adjustment was run through
completely and correctly, the instrument will be locked and hence
ready for operation.
Otherwise the instrument remains in the released and hence unsafe
condition.
Themainmenuisdividedintovesectionswiththefollowingfunc-
tions:
Setup: Settings, e.g. for measurement loop name, isotope, applica-
tion, background radiation, adjustment, signal output
Display: Settings, for example language, measured value display
Diagnosis: Information, for example, of device status, peak value,
simulation
Main menu