EasyManua.ls Logo

Vega VEGAPULS 62 - Measurement Errors with Constant Level; Constant Level Errors and Causes

Vega VEGAPULS 62
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
71
9 Diagnosis, asset management and service
VEGAPULS 62 • Probus PA
36505-EN-181127
1
2
Level
time
0
1 Real level
2 Level displayed by the sensor
Notes:
•
Wherever the sensor displays a constant value, the reason could
also be the fault setting of the current output to "Hold value"
•
If the level indication is too low, the reason could be a line resist-
ance that is too high
Measurement error with constant level
Fault description Cause Rectication
1. Measured value
shows a too low or
too high level
Level
time
0
•
Min./max. adjustment not correct
•
Adapt min./max. adjustment
•
Incorrect linearisation curve
•
Adapt linearisation curve
•
Installation in a bypass tube or standpipe,
hence running time error (small measure-
ment error close to 100 %/large error
close to 0 %)
•
Check parameter "Application" with
respect to vessel form, adapt if necessary
(bypass, standpipe, diameter)
2. Measured value
jumps towards 0 %
Level
time
0
•
Multiple echo (vessel top, product
surface) with amplitude higher than the
level echo
•
Check parameter "Application", especially
vessel top, type of medium, dished bot-
tom, high dielectric constant, and adapt if
necessary
3. Measured val-
ue jumps towards
100 %
Level
time
0
•
Due to the process, the amplitude of the
level echo sinks
•
A false signal suppression was not carried
out
•
Carry out a false signal suppression
•
Amplitude or position of a false signal has
changed (e.g. condensation, buildup);
false signal suppression no longer
matches actual conditions
•
Determine the reason for the changed
false signals, carry out false signal sup-
pression, e.g. with condensation

Table of Contents