18
VTK-TR-BA-099e-B
Quick Reference Guide MDV 3252-FH
6. Cleaning
Step 1: Pre-cleaning
Before placing the components into the beaker,
make sure that all parts are free from media.
If needed, push a cleaning rod or a fluidic brush
through the channels of all components that
were exposed to media (as shown with the
examples on the right).
Step 2: Clean all the components in an
ultrasonic bath
IMPORTANT NOTE!
Unless the solvent you use is compatible to the
O-ring/Tappet sealing, do NOT insert the O-
ring/Tappet Sealing into the solvent.
IMPORTANT NOTE!
Place tipped nozzle inserts (e.g. J-, C-, N27-type)
in a separate beaker. To avoid damage, do NOT
place tipped nozzle inserts together with other
components in one beaker.
Fill the beaker with appropriate solvent until all
parts are covered.
Put the beaker into an ultrasonic bath.