DGT / DGT60 / DGTQ / DGTP
16
YES enables the 50 Hz filter
Param. METRIC PARAMETERS
o StAbiL FILTER INTEGRATION
By pressing the PRINT key one accesses the selection of the type and degree of filter intervention for the
stability of the weight indication:
FLt 0 ÷ 3 Filter for simple weighing
F.F.100.1 ÷ 4 Filter for dosage at 100 Hz
h.r.0 ÷ 6 Filter for high resolution
dyn.0 ÷ 3 Filter for crane scale
F.F.50.1 ÷ 3 Filter for dosage at 50 Hz
F.F.200.1 ÷ 3 Filter for dosage at 200 Hz
F.F. 400 Filter for serial conversion/transmission at 400Hz
CuStoM Costumized filter for use of the manufacturer
The higher the filter value, and greater is its intervention relative to the type of filter used.
(!) FLt 3
(*) With approved instrument, one can select only the FLt 0, FLt 1, FLt 2, FLt 3 parameters
NOTE: The F.F.200.1 ÷ 3 and F.F.400 filters are not usable in the "DEP.CH" 2, 3 and 4 channel
functioning mode.
o (*) Auto-0 AUTO ZERO AT THE START UP
Automatic acquisition of the gross zero at the start-up (up to +/- 10% of the capacity):
- EnAb Enabled on scale 1
- diSAb Disabled
- CYCLE Executed cyclically on all the present scales. (§) This value is not visible if there is just one
scale (see nChAn parameter), or with more scales connected in a dependent way (see
tYPE parameter).
If the auto zero is enabled, the “C.PErC” message is displayed and one is aked to enter the value as a %
of the capacity of the autozero (included between 1 and 50)
(*) In case of approved instrument, the settable values are between 1 and 10.
(!) EnAb 50 (10 with approved instrument)
o (*) 0-PErC ZERO CAPACITY
This menu allows to set the % of the action capacity of the ZERO key. The settable values are in between 0
and 50. By entering the 0 value, the ZERO key is disabled.
(*) In case of approved instrument, the settable values are between 0 and 2.
(!) 50 (2 with approved instrument)
o (*) 0trACk ZERO TRACKING
This menu allows setting the zero tracking, in other words, the compensation parameter of the scale’s
thermal drift; the set value corresponds to the number of divisions that is reset in the fixed time of 1 second.
tr. ½ +/- half division.
tr. ¼ +/- one fourth of a division
tr. 1 +/- one division.
tr. 2 +/- two divisions.
tr. no tracking disabled.
(!) tr. ½
(*) In case of approved instrument, it’s possible to select only the tr. no, tr. ½ and tr. ¼ parameters