6.2 Gas Measurement Process – Best Practice
Depending on preferences the exact reading procedure can change.
The following method is considered best practice and when followed correctly will allow
quick and consistent readings to be recorded.
1) When the instrument is first switched on it should be purged with fresh air and
allowed to stabilise for a few minutes.
2) At this point it is good practice to verify the CO
2
channel. Calibrate only if required.
This option is available via the ‘Calibration’ menu.
The instrument is now ready to take the first reading.
3) Connect the sample tube, if relevant (always use the sample filter) from the sample
point to the inlet port of the instrument, ensuring the filter is seated correctly.
4) Press the ‘Pump’ key to draw a sample into the instrument.
Notice the main gas readings start to change. It is recommended to run the pump
until the gas readings have stabilized (approx. 30 seconds) then press the soft-key
‘Store’.
5) The pump will stop and the operator will be prompted to enter an ID code to identify
the reading. A “reading stored” confirmation message will be displayed briefly before
returning to the ‘Main Read Screen’.
Note: The ID code prompt can be switched on or off. This option is accessed via
the ‘Settings’ menu, then press ‘Key 1 - Prompt for ID: Yes or No’. If the
ID prompt is set to ‘No’ the reading is stored using the default logging ID.