EasyManua.ls Logo

VICI M50 - Page 15

VICI M50
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11
1. IfthescreeninFigure 9 is not already open, go to the main menu, and
choose Utility > Test.
2. IfthetubingIDis0.010”orless,enter
FlowRate: 200µL/min
FixedVolume: 1000µL
Click the Enable box
IfthetubingIDislargerthan0.010”,enter
FlowRate: 9000µL/min
FixedVolume: 1000µL
Click the Enable box
(In the Test Mode, user-entered flow parameters cannot be saved; when you
exit the program, these values will revert to defaults.)
NOTE:SmallerIDandlongerlengthtubingcancauseexcessive
restriction. Adjust flow rates accordingly. Maximum backpressure
is100psi.
3. ClicktheDispensebutton.Thedesignatedvolumeofisopropylalcohol
should move through the tubing and into the receiver bottle, removing
any air bubbles and priming the system. If it does, the pump is primed
and ready for use.
NOTE: If a buffer will be used, flush all the isopropyl alcohol out of
the lines by placing the inlet tube in a vessel of water, changing the
xedvolumeto2000µL,andclickingtheDispensebutton.
4. If the pump doesn’t seem to draw any isopropyl alcohol into the tubing:
(a) Check for leaks around the fittings and make sure the tubing is well
seated in the pump head.
(b) Make sure the end of the tubing is free to draw liquid into the pump.
(TIP:Cutthetipendthatgoesintotheisopropyl
alcohol reservoir at a 45° angle.)
(c) Touch the pump to make sure it is on.
Ifthepumpstilldoesn’tseemtofunctionproperly,consultthefactory.
Figure 9:PrimingthepumpintheTestMode(M6valuesshown)
CLICKTHEBOX
FOR A ONE TIME
FIXEDVOLUME
ASPIRATEOR
DISPENSE.
LEAVETHEBOX
UNCLICKEDFOR
CONTINUOUS
PUMPINGATTHE
ENTEREDFLOW
RATE.