40mA ±(2.0%+5dgts) 0.01mA 40 to 400Hz
400mA ±(2.0%+5dgts) 0.1mA 40 to 400Hz
4A ±(2.5%+5dgts) 0.001A 40 to 400Hz
20A ±(2.5%+5dgts) 0.01A 40 to 400Hz
Resistance
400 ±(1.2%+2dgts) 0.1
4k ±(1.0%+2dgts) 0.001k
40k ±(1.0%+2dgts) 0.01k
400k ±(1.0%+2dgts) 0.1k
4M ±(1.2%+2dgts) 0.001M
40M ±(1.5%+2dgts) 0.01M
Capacity
40nF ±(3.0%+10dgts) 0.01nF
400nF ±(3.0%+5dgts) 0.1nF
4µF ±(3.0%+5dgts) 0.001µF
40µF ±(3.0%+5dgts) 0.01µF
100µF ±(4.0%+5dgts) 0.1µF
Continuity tester
An audio signal of a continuity is generated for resistance values
less than approximately 70. Moreover, for values up to 100 the
indication is only given in "" without an audio signal.
Diode test
The maximum measuring voltage at the object is about 1.48VDC
(maximum), display resolution 1mV.
Frequency measurement / pulse-pause-ratio / pulse duration mea-
surement
10Hz – ±(0.1%+3dgts) 0.01Hz –
10MHz 10kHz
input sensitivity 0.3Veff from 10Hz to 1MHz
Input sensitivity 0.6Veff > 1MHz
46

Related product manuals