EasyManua.ls Logo

Vyaire VELA - Chapter 5: One-Year Preventive Maintenance; Contents of Annual PM Kit; General Instructions and Warnings for PM; Required Tools for Annual PM

Vyaire VELA
111 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
viii
33373-001 Version A
Anti-static or electrically conductive hoses or tubing should not be used within the patient circuit.
If a mechanical or electrical problem is recognized while running the Operational Verification
Tests, or while operating the ventilator, the ventilator must be removed from use and referred to
qualified personnel for servicing. Using an inoperative ventilator may result in patient injury.
When a low gas supply alarm occurs, the oxygen concentration delivered to the patient will
differ from that set on the O2 control setting.
A source gas failure will change the FIO2 and may result in patient injury.
The functioning of this equipment may be adversely affected by the operation of other
equipment nearby, such as high frequency surgical (diathermy) equipment, defibrillators, short-
wave therapy equipment, “walkie-talkies,” or cellular phones.
Do not block or restrict the Oxygen bleed port located on the instrument back panel. Equipment
malfunction may result.
Electric shock hazard - Do not remove any of the ventilator covers or panels. Refer all servicing
to an authorized Vyaire, service technician.
A protective ground connection by way of the grounding conductor in the power cord is essential
for safe operation. Upon loss of protective ground, all conductive parts including knobs and
controls that may appear to be insulated, can render an electric shock. To avoid electrical
shock, plug the power cord into a properly wired receptacle, use only the power cord supplied
with the ventilator, and make sure the power cord is in good condition.
Cautions
The following cautions apply any time you work with the ventilator.
When replacing fuses, ensure that new fuses are of the same type and value as those being
replaced. Incorrect fuses can cause damage to the ventilator.
A battery that is fully drained (i.e. void of any charge) may cause damage to the ventilator and
should be replaced.
All accessory equipment that is connected to the ventilator must comply with
CSA/IEC601/UL2601.
To avoid damage to the equipment, clean the air filter regularly.
The following cautions apply when cleaning the ventilator or when sterilizing ventilator
accessories.
Do not sterilize the ventilator. The internal components are not compatible with sterilization
techniques.
Do not gas sterilize or steam autoclave tubing adapters or connectors in place. The tubing will,
over time, take the shape of the adapter, causing poor connection and possible leaks.
DO NOT submerge the ventilator or pour cleaning liquids over or into the ventilator.

Table of Contents

Other manuals for Vyaire VELA

Related product manuals