27
The available sterilization cycles
For your safety and the safety of your patients
Never exceed the maximum load weight limits as specified in the CYCLE PROGRAM TABLE (see next page) as this
could impair the sterilization process.
The sterilizer is not suitable for the sterilization of liquids, textiles and hollow instruments .
In total there are four sterilization cycles available.
To select the appropriate cycle
for your load and requirements: see CYCLE SELECTION GUIDE in the table below.
For more details
about the cycle parameters: see the CYCLE PROGRAM TABLE in the next page.
CYCLE SELECTION GUIDE
CYCLE NAME DESCRIPTION REF. STANDARD
Very fast processing of unwrapped dental instruments
ANSI/AAMI ST55
Wrapped dental instruments (2 kg).
Cycle for dental instruments, unwrapped (4 kg) or single wrapped (1 kg) , withstanding the
temperature of 134 degrees.
EN 13060
(Type S)
Fast processing of a limited load (2 kg) of dental instruments , not bagged.
NOTE The weight in the table above is to be intended as the maximum permitted for the cycle.