41
6
Cleaning the chamber filter
Empty the sterilizer chamber by removing the trays and the rack.
1–2: Remove the filter cap at the back of the chamber (bottom/center)
by turning it counter-clockwise.
3: Remove the cartridge filter and rinse it with tap water.
4–5–6: Insert the filter in the cap, attach the filter cap and lock it by
turning clockwise.
Cleaning the external surfaces of the sterilizer
Clean all external sterilizer covers with a slightly damp cloth moistened with water.
For better cleaning results, clean with W&H MC-1000 cleaning solution.
NOTICE: Never use any other disinfectant, detergent or abrasive product, as they might result aggressive for the external
covers and damage them.
Monthly or 50-cycle maintenance