W&H MS - Periodic Filter Replacement; Bacteriological and Dust Filter Replacement

W&H MS
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42
3 month or 400-cycle maintenance
Replacing the bacteriological filter
Open the service door.
Unscrew the bacteriological filter by hand (counter-clockwise).
Screw on the new bacteriological filter (clockwise) and tighten it snug.
Replacing the dust filter
Pull out the dust filter from underneath the sterilizer.
Detach the used filter from the handle.
Attach the new filter to the handle.
Slide the filter back into its original position.
Remember to reset the counter after replacement (see following pages).
Remember to reset the counter after replacement (see following pages).