4000 cycle/5 years general check and service
REPLACEMENT PARTS CLEANING CHECKS
Solenoid valves
Door locking system
(every 20000 cycles only)
Sterilization chamber and external surfaces Pneumatic connections
Chamber filter Electrical connections
Internal cleaning, with particular care for the
condenser fins and the main board
Temperature and pressure calibration
Door locking system
Pressure safety valve
Steam diffuser plate Safety systems
Regular service is imperative to ensure continuous and effective operation of the sterilizer.
It is recommended to carry out a general service every 4000 cycles or five years by an authorized service technician.
The service includes replacement of consumables and other important internal components, a check of the entire unit
with special care for the safety systems, and cleaning of areas and components that cannot be accessed by the user.