Alarms
NOTE: for any alarm not listed in this table, call technical service.
Alarm code DESCRIPTION ACTION
E010
Power failure during a cycle Load cannot be considered sterile. Repeat the cycle.
E02x
Internal voltage error Switch the sterilizer OFF and ON. If the problem persists call service.
E041
Cycle counter lost
Switch the sterilizer OFF and ON. If the problem persists call service.
NOTE: Initiating a sterilization cycle is still possible.
E042
Internal clock error
Set date and time - Switch the sterilizer OFF and ON. If the problem persists call service.
NOTE: Initiating a sterilization cycle is still possible.
E060
Internal voltage error
Disconnect optional accessories from 24VDC rear plug - switch the sterilizer OFF and ON. If the problem
persists call service.
E080
Internal overheating Check the dust filter and ensure that the sterilizer fan is not blocked.
E090
Internal voltage error Switch the sterilizer OFF and ON. If the problem persists call service.
E100
Phase timeout Check water level in the clean water tank. Reset the thermal overload. If the problem persists call service.
E101
Internal probe error Switch the sterilizer OFF and ON. If the problem persists call service.
E104
Phase timeout Check water level in the clean water tank. Reset the thermal overload. If the problem persists call service.
E121
Internal probe error Switch the sterilizer OFF and ON. If the problem persists call service.
E130
Overpressure during the sterilization phase
Clean the chamber from residuals of detergents, disinfectants and other chemicals.
Replace the clean water if it is suspected to be contaminated with chemicals.
Ensure all the load is clean rinsed and free from any chemicals before sterilizing.
Repeat the cycle. If the problem persists call service.
E131
Temperature fluctuation during the steril. phase
E132
Temperature difference too high during PG1
E140
Low pressure during the sterilization phase
E150
Low temperature during the sterilization phase
E160
Overtemperature during the sterilization phase
E163
Overpressure detected
E164
Overpressure during PG1