Alarms
NOTE: for any alarm not listed in this table, call technical service.
Alarm code DESCRIPTION ACTION
E180-E181
Internal probe error Switch the sterilizer OFF and ON. If the problem persists call service.
E183
Pressure drop timeout Clean the chamber filter. If the problem persists call service.
E184
Overtemperature detecded If the problem persists call service.
E215
Fan blocked or faulty electronic control Call service.
E230
Internal probe error Switch the sterilizer OFF and ON. If the problem persists call service.
E231
Overtemperature detecded If the problem persists call service.
E232-E233-E234
Internal probe error Switch the sterilizer OFF and ON. If the problem persists call service.
E240
Heating element error Wait for the chamber to cool down. Reset the thermal overload. If the problem persists call service.
E241
Heating element overheating
Switch the sterilizer OFF. Wait for the chamber to cool down. Switch the sterilizer ON. If the problem
persists call service.
E242
Chamber filter blocked Clean the chamber filter. If the problem persists call service.
E310-E320-E380-
E390
Vacuum timeout
Check the door seal; clean or replace if necessary. Clean the chamber face side. Clean the chamber
filter. If the problem persists call service.
E510
Door motor: failure after cycle completion Switch the sterilizer OFF and ON. If the problem persists call service.
E520
Door motor: locking timeout If the problem persists call service.
E570
Door motor: unable to detect the door position Switch the sterilizer OFF and ON. If the problem persists call service.
E580
Door motor: door locked check signal lost If the problem persists call service.
E59x
Door motor error Switch the sterilizer OFF and ON. If the problem persists call service.
E950
Internal memory error
Switch the sterilizer OFF and ON. If the problem persists call service.
NOTE: Initiating a sterilization cycle is still possible.
E95x-E96x
Internal memory error Switch the sterilizer OFF and ON. If the problem persists call service.
E990
Manual stop The cycle has been aborted by the user. Re-process the load.