B Specifications
The reproducibility of the specifications presented in this document depends on the conditions in
individual laboratories. Contact the Waters Technical Service organization for additional
information about specifications.
B.1 CH-A physical specifications
Attribute Specification
Height 7.6 cm (3 inches)
Width 34.3 cm (13.5 inches)
Depth (without drip tray fitting) 62.9 cm (24.8 inches)
Weight 5.7 kg (12.5 lb)
Access to APH assembly, column in-line filter,
column, and column-outlet tubing
Inside column compartment via access door
Number of columns One, 150-mm column, maximum, with guard
One, 150-mm column, maximum, with column
in-line filter
Column sizes 4.6-mm ID, maximum
Column guards or in-line filters, 20-mm length,
maximum
B.2
CH-A environmental specifications
Attribute Specification
Ambient operating temperature 4 to 40 °C
Ambient operating humidity 20 to 80%, noncondensing
Ambient shipping and storage temperature -30 to 60 °C
Ambient shipping and storage humidity 20 to 85%, noncondensing
December 4, 2019, 715005052 Version 02 (previously released as Rev.B)