B.3 CH-A electrical specifications
Attribute Specification
Pollution degree
a
2
Moisture protection
b
Normal (IPXO)
Supply voltage, nominal 24 Vdc
Line frequency 0 Hz
Power consumption 84 W
a. Pollution Degree 2 – A measure of pollution on electrical circuits that can produce a reduction of dielectric strength or
surface resistivity. Degree 2 refers only to normally nonconductive pollution. Occasionally, however, expect a temporary
conductivity caused by condensation.
b. Moisture Protection – Normal (IPX0) – IPX0 means that no Ingress Protection against any type of dripping or sprayed water
exists. The “X” is a placeholder that identifies protection against dust, if applicable.
B.4 CH-A input/output specifications
Attribute Specification
25-pin, D-series sub-connector Control and communication from the sample
manager.
Also serves as a power interface for the CH-A.
B.5
CH-A wetted materials of construction
Component Material
Tubing and fittings, APH assembly 316 SST, PEEK , MP35N (Bio)
Column outlet tubing and fittings 316 SST, PEEK
Drip tray Polyethylene
Drain tubing Tygon 2475
December 4, 2019, 715005052 Version 02 (previously released as Rev.B)