Table 1–1: Supported column configurations
Quantity Description
1 Column length = 150 mm; column ID = 4.6
mm, maximum;
Column guard or column in-line filter = 20 mm,
maximum;
1.4
eCord technology
ACQUITY UPLC columns include an eCord column chip that tracks the usage history of the
column. The eCord column chip interacts with the system software, recording information for as
many as 50 sample queues run on the column. In regulated environments, the eCord column chip
provides documentation of the column used in the validation method. The eCord column chip
provides documentation of the column used for each chromatographic run and records the
following information:
• The name of the sample set (or sample list) run on the column.
• Number of injections onto the column.
• Number of samples injected onto the column.
• The highest pressure that the column has experienced (and the date).
• The highest temperature the column has experienced (and the date).
In addition to the variable column usage data, the eCord column chip also stores fixed column
manufacturing data, including:
• unique column identification.
• certificate of analysis.
• QC test data.
When you attach the column's eCord fob to the receptacle on the column compartment, the chip
automatically records and stores system information. You need take no further action.