10. Roll up a nonabrasive, lint-free wipe and use it to absorb the liquid from the leak-sensor
reservoir and its surrounding area.
11. With a cotton swab, absorb any remaining liquid from the corners of the leak-sensor
reservoir.
12. Align the sensor's T-bar with the slot in the side of the leak-sensor reservoir and slide the
leak sensor into place.
13. In the console, select Sample Manager from the system tree.
14. In the Sample Manager information window, click Control > Reset module to reset the
sample manager.
December 4, 2019, 715005052 Version 02 (previously released as Rev.B)