Figure 3–6: Removing the APH assembly
APH assembly
APH retainer clip
4. Disconnect the column outlet fitting.
5. Loosen the #2 locking cap nut on the APH assembly before removing the column
assembly.
6. Place the column gripping tool onto the column.
7. While holding the column with the column gripping tool, loosen the #1 knurled compression
fitting and remove the column assembly.
Figure 3–7: Loosening the column from the compression fitting
December 4, 2019, 715005052 Version 02 (previously released as Rev.B)
Page 76