3 Maintenance Procedures
82 September 26, 2013, 715004208 Rev. A
2. Unscrew the finger-tight fitting and remove the old ISM solvent filter,
ferrule, and fitting from the solvent tubing.
3. Separate the new filter assembly and place the new finger-tight fitting
and ferrule onto the solvent tubing.
4. Insert the solvent tubing into the filter, ensuring that it bottoms out in
the filter.
5. Screw the finger-tight fitting and ferrule into the filter to secure the
solvent tubing in place.
6. Insert the filter end of the solvent tubing into the solvent bottle.
7. Shake the filter, to remove any air from it.
8. Prime the ISM (see page 67).
Finger-tight
fitting
Ferrule
Filter
Finger-tight
fitting
Ferrule
Filter