Replacing a restrictor module in the optional ACQUITY UPLC or Alliance Flow-splitter kit
September 26, 2013, 715004208 Rev. A 87
Replacing a restrictor module in the optional ACQUITY
UPLC or Alliance Flow-splitter kit
Required materials
• Gloves: clean, powder-free, chemical-resistant
• Restrictor module
Required tool
T10 TORX driver (startup kit)
To replace a restrictor module:
1. Disconnect the inlet 1, outlet 1, inlet 2, and outlet 2 tubing from the
restrictor module that you are replacing.
Restrictor module (3)
Inlet 1:
From
column
Outlet 1:
To optical
detector
Inlet 2:
From ISM
outlet
Outlet 2:
To QDa