EasyManua.ls Logo

Wavetronix smartsensor advance - Internet Connection

Default Icon
136 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CHAPTER 4  COMMUNICATION 41
Figure 4.2 – Serial Connection (left) and Advanced Settings (right)
e Advanced Settings screen contains the following elds:
Timeout (ms) – Allows you to set an additional amount of time (in milliseconds) that
SSMA will wait for a response when communicating with the sensor.
Buer (bytes) – Contains the number of bytes used by SSMA to store data received
from the sensor.
Parity – is should always be set to None.
Stop Bits – is should always be set to 1.
Data Bits – is should always be set to 8.
Note
When a connection is made using the Isolated Sensor tab instead of the Multi-drop
Network tab, the software will connect to the first sensor it finds. If you have multi-
ple sensors and want to be able to connect to each one, use the Multi-drop Network
tab (see Multi-drop Network section below).
Internet Connection
1 Click on Communication.
2 Select the Internet tab (see Figure 4.3).
3 Enter the IP address of the Serial-to-Ethernet adapter to which the sensor of interest
is connected, or the domain name associated with the sensor in the Address eld. e
IP address is four octets, each ranging from 0-255, separated by dots (“.”). An URL
address can also be entered into this eld (i.e. sensor.earlink.com).
4 Enter the port number assigned to the Ethernet-to-Serial adapter in the Port eld.
is will be an integer value in the range of 0-65536. If you are using a Click 301
device, the default port number will be 10001. Click the Advanced… button for ad-

Other manuals for Wavetronix smartsensor advance

Related product manuals