calibration weights will be entered in this unit. The units parameter is both sealed and
audited.
9.2 “Capacity”
Capacity sets the maximum capacity of the indicator, in setup units. This parameter
is both sealed and audited. Factory default is 0, which must be changed before the
indicator will weigh.
9.3 “Resltn”
Parameter that sets the resolution of the indicator. Resolution is limited to values
available on the scroll list. Resolution is set in terms of the setup units. This parameter
is both sealed and audited.
9.4 “Stability”
This parameter controls how many consecutive weight readings are required to be within
the motion sense band before the weight indication is considered to be stable. The
indicator reads the analog input 7.5 Hz (7.5 times per second), so the default setting
of four requires about a half second of stable weight. Either the net or gross light will
come on when the weight is stable. This parameter is both sealed and audited.
9.5 “Motion sns”
Amount of motion, in divisions, allowed before the weight is considered unstable.
Default is one division. This parameter is both sealed and audited.
9.6 “Prefilter”
Length of the prefilter buffer. Larger numbers provide slower and cleaner weight
readings. Default is 2. This parameter is both sealed and audited. Range?
9.7 “AZT”
Auto zero tracking on/off. This parameter is neither sealed nor audited. When on, stable
weights within the AZT band” of zero will automatically rezero the scale.
9.8 “AZT band”
Amount of weight, in divisions, that can be automatically zeroed out at one time. Default
is 1 division. Parameter is sealed and audited.
9.9 “Calibrate”
This function starts the indicator calibration routine. It is sealed and audited. Refer to
the calibration section of this manual for details.