CP 20_______________________________________________________________________Section 3
Summary of Configuration Menus
Configuration
Parameter
Definition Configuration Screen
Software Version The firmware version of the unit Screen One
Cart Number User-defined (up to 4 digits) Screen One
Site Number User-defined (up to 4 digits) Screen One
Site Name User-defined (up to 30 alphanumeric characters) Screen One
Telephone Number User-defined (up to 30 alphanumeric characters) Screen One
Volume Range from 0 – 8 Screen One
Battery Timeout 10min, 30min, 60min Screen One
AC Filter None, 60Hz, 50Hz Screen Two
Filter 40Hz, 150Hz, 300Hz Screen Two
Paper Speed 25 mm/s, 50mm/s Screen Two
ID Format Short, Long, Long 2, Standard Screen Two
Height/Weight Units lb/in or kg/cm Screen Two
Date Format US (mm/dd/yy) or European (dd.mm.yy) Screen Two
Plot Format Channels printed: 3+1, 6, 3+3, 12, 6+6 Screen Two
Interpretation Yes or No Screen Two
Reasons Yes or No Screen Two
Append to ECG Unconfirmed, Reviewed by, Blank Screen Two
#of copies 0 - 9 Screen Three