Where the integrity of external protective earth conductor arrangement is in doubt, the CP 20 shall be operated
from its internal electrical power source.
All signal input and output (I/O) connectors are intended for connection of only devices complying with IEC
60601-1, or other IEC standards (e.g. IEC 60950), as appropriate to the device. Connecting additional devices
to the CP 20 may increase chassis and/or patient leakage currents. To maintain operator and patient safety,
consideration should be given to the requirements of IEC 60601-1-1, and leakage currents should be measured
to confirm no electric shock hazard exists.
The CP 20 has not been designed for use with high-frequency (HF) surgical equipment and does not provide a
protective means against hazards to the patient.
The quality of the signal produced by the electrocardiograph may be adversely affected by the use of other
medical equipment, including but not limited to defibrillators and ultrasound machines.