Operating Temperature:
15O
to
35°C
(59O
to 95°F). All
specifications apply over this temperature range.
Storage Temperature:
-20”
to +
40°C
(-4O
to +
104OF).
Printer Paper:
Thermal type, 58mm (2.28”) wide x
19.8m
(65’) long
with end-of-roll indicator, Welch Allyn No. 53600. Must be stored in
dark, cool location, to prevent exposure of paper. 205 double-ear
printouts per roll.
Battery:
7.2V
nickel-cadmium rechargeable, Welch Allyn No. 72700,
approximately 60 minutes continuous use. (See pg. 24)
Tips:
Four color-coded sizes.
Probe Tone:
Frequency: 226 Hz
+3%
Amplitude: 85
f
3
dB
re 20
uPa
in an ANSI HA-1
(2cc)
coupler.
Distortion: 5% maximum total harmonic distortion in an ANSI
HA-1
(2cc)
coupler.
Pump:
Direction of sweep: positive to negative pressure.
Speed: 200
rt
20
daPa/second
average during data acquisition
period
Pressure Measurement System:
Range:
+
200
to
-300
daPa
Display resolution: 20
daPa
Accuracy:
f
15% or
+
10
daPa,
whichever is greater.
Compensation: Auto-zero with each test cycle
Admittance Measurement System:
Range: 0.2 to 3.8 mmho total
0 to 1.5 mmho for Ya
0.2 to
2cc
for + 200 Vea
Display resolution:
.l
mmho for Ya
0.2~~
for
+
200 Vea
Accuracy:
kO.1
mmho or
f
5%,
whichever is greater.
45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn MicroTymp and is the answer not in the manual?

Welch Allyn MicroTymp Specifications

General IconGeneral
Measurement MethodInfrared
Battery LifeApproximately 1000 measurements
TypeTympanic (ear) thermometer
DisplayLCD
MeasurementBody temperature
MemoryLast reading memory
Battery Type2 AAA batteries

Related product manuals