EasyManuals Logo

Welch Allyn SureTemp Plus 690 Series User Manual

Welch Allyn SureTemp Plus 690 Series
34 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
22 Welch Allyn SureTemp Plus
Emissions and Immunity Information
Electromagnetic Emissions
The Welch Allyn SureTemp Plus is intended for use in the electromagnetic environment specified below. The customer or the user of the Welch
Allyn thermometer should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1 The SureTemp Plus uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B The SureTemp Plus is suitable for use in all establishments other than domestic, and
those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposespurposes, provided the following warning is
heeded:
WARNING: This equipment/system is intended for use by healthcare
professionals only. This equipment/system may cause radio interference or may disrupt
the operation of nearby equipment. It may be necessary to take mitigation measures,
such as re-orienting or relocating the device or shielding the location.
Harmonic emissions
IEC 61000-3-2
Class B
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn SureTemp Plus 690 Series and is the answer not in the manual?

Welch Allyn SureTemp Plus 690 Series Specifications

General IconGeneral
BrandWelch Allyn
ModelSureTemp Plus 690 Series
CategoryThermometer
LanguageEnglish

Related product manuals